Notifications
You must be signed in to change notification settings - Fork 1.4k
Open
Description
Describe the bug
There is no limit on the amount of non-pump insulin accepted by Loop.
Attach an Issue Report
N/A
To Reproduce
Steps to reproduce the behavior:
- Tap on Insulin Delivery Graph
- Tap on '+' at upper right
- Enter a bolus amount, in this case 1000 U
- Tap on log dose
- Loop accepts that amount
This can also happen by entering a value in Apple Health under insulin when it was supposed to be a carb entry or glucose entry.
Expected behavior
I think this is an oversight in the design.
In my opinion, the should be limited - I suggest something like 2 or 3 times the maxBolus amount.
Screenshots
- The screenshot on the left is after entering 1000 U of external insulin in Loop.
- The screenshot on the right is after entering 300 U in Apple Health.
Phone
- Hardware: SE 3rd gen
- OS Version: iOS 26.0
Loop Version
- v3.8.0, but this has been there for a while
CGM
- Irrelevant
Pump
- Irrelevant
Additional context
Add any other context about the problem here.
Metadata
Metadata
Assignees
Labels